1.
Resetting to factory default clears ALL user-entered settings, including sensor and
analog output calibration, and returns ALL settings and calibration values to the
factory defaults.
2.
Resetting a sensor calibration to the default value clears user-entered calibration
data for the selected sensor but leaves all other user-entered data unaffected.
3.
Resetting the analog output calibration clears only the user-entered analog output
calibration. It leaves all other user-entered settings unchanged.
7.8.2
Procedure - resetting the transmitter
Complete the following steps to reset the transmitter.
1.
Press MENU.
The main Menu screen appears.
2.
Move to Program and press ENTER.
3.
Scroll to the bottom of the screen and continue scrolling until
Reset Analyzer
is
highlighted. Press ENTER.
4.
Choose whether to reset all user-entered values (
Factory Defaults
), sensor calibration
(
Sensor Cal Only
), or output calibration (
Output Cal Only
).
If you choose
Sensor Cal Only
or
Output Cal Only
, a second screen appears in which you
can select which sensor or output calibration to reset.
5.
To return to the main display, press MENU and then EXIT.
Programming the transmitter
Instruction Manual
57
Summary of Contents for Rosemount TCL
Page 4: ......
Page 8: ...Contents iv Rosemount TCL ...
Page 16: ...Description and specifications 8 Rosemount TCL ...
Page 18: ...Principles of operation 10 Rosemount TCL ...
Page 26: ...Installation 18 Rosemount TCL ...
Page 34: ...Wiring 26 Rosemount TCL ...
Page 46: ...Display and operation 38 Rosemount TCL ...
Page 66: ...Programming the transmitter 58 Rosemount TCL ...
Page 68: ...Digital communications 60 Rosemount TCL ...
Page 106: ...Troubleshooting 98 Rosemount TCL ...