The transmitter will not update the zero current. For troubleshooting assistance, see
10.
To return to the main display, press MENU and then EXIT.
9.3.3
Procedure - calibrating the sensor
Complete the following steps to calibrate the total chlorine sensor with the full scale
standard solution.
1.
If the sensor was just zeroed, place the sensor back in the flow cell. Confirm that
excess sample is flowing down the inside tube of the overflow sampler. Also, verify
that reagent is being delivered to the sample and that the air pump is working.
2.
Adjust the chlorine concentration until it is near the upper end of the operating
range. Wait until the transmitter reading is stable before starting calibration.
When the TCL is first started up or when a new sensor is put in service, allow at least
30 minutes for the reading to stabilize.
3.
Press MENU.
The main Menu screen appears. The cursor is on
Calibrate
.
4.
Press ENTER.
5.
Choose the sensor you wish to calibrate.
Sensor 2
appears only if you have a dual input Rosemount 1056 Transmitter.
6.
Choose
Total Chlorine
.
Calibration
Instruction Manual
67
Summary of Contents for Rosemount TCL
Page 4: ......
Page 8: ...Contents iv Rosemount TCL ...
Page 16: ...Description and specifications 8 Rosemount TCL ...
Page 18: ...Principles of operation 10 Rosemount TCL ...
Page 26: ...Installation 18 Rosemount TCL ...
Page 34: ...Wiring 26 Rosemount TCL ...
Page 46: ...Display and operation 38 Rosemount TCL ...
Page 66: ...Programming the transmitter 58 Rosemount TCL ...
Page 68: ...Digital communications 60 Rosemount TCL ...
Page 106: ...Troubleshooting 98 Rosemount TCL ...