Replacing the air pump
Complete the following steps to replace the air pump on your total chlorine system.
WARNING!
HAZARDOUS VOLTAGE
Can cause severe injury or death. Disconnect power before servicing.
Procedure
1.
Disconnect power to the transmitter.
2.
Refer to
. Disconnect the reagent and air injection tubes. Disconnect the
suction and discharge tubing by turning the Luer fitting in the direction shown in the
figure. Disconnect the air pump inlet tubing from the barbed tubing in the bottom
of the enclosure.
Replacing the Air Pump
Figure 10-12:
3.
Remove the four screws (circled in
and
pump access panel. Pull out the pump and panel.
Maintenance
80
Rosemount TCL