For best EMI/RFI protection, use shielded output signal cable enclosed in earth-grounded
metal conduit.
Keep output signal wiring separate from power wiring. Do not run signal and power or
relay wiring in the same conduit or close together in a cable tray.
4.3.3
Alarm wiring
The alarm relay terminal strip is located just below the power connector on the power
supply board.
See
Alarm relay connections
Figure 4-3:
Keep alarm relay wiring separate from signal wiring. Do not run signal and power or relay
wiring in the same conduit or close together in a cable tray.
4.4
Sensor wiring
If it is necessary to replace the sensor cable, refer to the instructions below.
1.
Shut off power to the transmitter.
Wiring
22
Rosemount TCL