B
Removal or repositioning of the magnetic
fixture
Prerequisites
•
Ensure you have 4 x protective caps for the magnetic fixture and the
protective cap for the sensor.
•
One person needs to hold the sensor in position while another follows
the procedure.
Procedure
1. Remove the 4 sensor securing bolts, washers, and nuts as displayed
, then place the protective cap back on the sensor and
set it aside.
Figure B-1: Magnetic fixture
A. Frame
B. Ball stud
C. Protective cap
D. Sensor securing bolts
E. Magnetic unit
2. Some one must hold the magnetic fixture in position.
3. Slide a flathead screwdriver under one of the magnetic units as
displayed in
.
January 2022