6.6
Dangerous goods regulation
The magnets in the sensor and magnetic fixture are shielded for
transportation and meet the IATA Dangerous Goods Regulations for
magnetic fields. The sensors are safe for air transportation.
6.7
USA
6.7.1
I5 USA Intrinsic Safety (IS)
Certificate:
SGSNA/17/SUW/00281
Standards:
UL 913 — 8th Edition, Revision Dec 6 2013
Markings:
CLASS I, DIV 1, GP ABCD, T4…T2, Tamb = –50 °C to +75 °C,
IP67
6.8
Canada
6.8.1
I6 Canada Intrinsically Safe (IS)
Certificate:
SGSNA/17/SUW/00281
Standards:
CAN/CSA C22.2 No. 157-92 (R2012) +Upd1 +Upd2
Markings:
CLASS I, DIV 1, GP ABCD, T4…T2, Tamb = –50 °C to +75 °C,
IP67
6.9
Europe
6.9.1
I1 ATEX Intrinsic Safety
Certificate:
Baseefa17ATEX062X
Standards:
EN IEC 60079-0:2018
EN 60079-11: 2012
Markings:
II 1 G, Ex ia IIC T4…T2 Ga, Tamb = –50 °C to +75 °C, IP67
Special Condition for Safe Use (X):
1. The plastic mounting foot may present a potential electrostatic
ignition risk and must not the rubbed or cleaned with a dry cloth.
2. The equipment may be attached to process pipework at a
temperature of up to 200 °C as follows:
a. –50 °C ≤ Ta ≤ +120 °C for T4
b. –50 °C ≤ Ta ≤ +190 °C for T3
c. –50 °C ≤ Ta ≤ +200 °C for T2
January 2022
Quick Start Guide
Quick Start Guide
39
Summary of Contents for Rosemount ET310
Page 4: ...Quick Start Guide January 2022 4 Emerson com Rosemount ...
Page 42: ...6 14 China RoHS Quick Start Guide January 2022 42 Emerson com Rosemount ...
Page 44: ...Quick Start Guide January 2022 44 Emerson com Rosemount ...
Page 50: ...Quick Start Guide January 2022 50 Emerson com Rosemount ...
Page 51: ...January 2022 Quick Start Guide Quick Start Guide 51 ...