62
Reference Manual
00809-0100-4697, Rev GA
Appendix A: Reference Data
November 2014
Reference Data
Table A-2. Analog Sensors
A.4.9
Ambient temperature effect
Transmitter may be installed in locations where the ambient temperature is between -40 and
85 °C (-40 and 185 °F).
A.4.10
Ambient temperature effects
Sensor option
Alarm levels
Accuracy
4–20mA (Rosemount Standard)
3.9 to 20.8 mA
± 0.01mA
4–20mA (NAMUR)
3.8 to 20.5 mA
± 0.01mA
NIST type
Accuracy per 1.0 °C (1.8 °F) change in ambient temperature
(1)
°
C
Temperature range (°C)
RTD
Pt 50 (
α
= 0.00391)
• 0.004 °C (0.0072 °F)
N/A
Pt 100 (
α
= 0.00391)
• 0.002 °C (0.0036 °F)
N/A
Pt 100 (
α
= 0.00385)
• 0.003 °C (0.0054 °F)
N/A
Pt 100 (
α
= 0.003916)
• 0.003 °C (0.0054 °F)
N/A
Pt 200 (
α
= 0.003916)
• 0.004 °C (0.0072 °F)
N/A
Pt 200 (
α
= 0.00385)
• 0.004 °C (0.0072 °F)
N/A
Pt 500
• 0.003 °C (0.0054 °F)
N/A
Pt 1000
• 0.003 °C (0.0054 °F)
N/A
Cu 10
• 0.03 °C (0.054 °F)
N/A
Cu 100 (a=428)
• 0.002 °C (0.0036 °F)
N/A
Cu 50 (a=428)
• 0.004 °C (.0072 °F)
N/A
Cu 100 (a=426)
• 0.002 °C (0.0036 °F)
N/A
Cu 50 (a=426)
• 0.004 °C (.0072 °F)
N/A
Ni 120
• 0.003 °C (0.0054 °F)
N/A
Thermocouple (R = the value of the reading)
Type B
• 0.014 °C
• 0.032 °C - (0.0025% of (R - 300))
• 0.054 °C - (0.011% of (R - 100))
• R
≥
1000
• 300
≤
R < 1000
• 100
≤
R < 300
Type E
• 0.005 °C + (0.00043% of R)
• All
Type J, DIN Type L
• 0.0054 °C + (0.00029% of R)
• 0.0054 °C + (0.0025% of |R|)
• R
≥
0
• R
<
0
Type K
• 0.0061 °C + (0.00054% of R)
• 0.0061 °C + (0.0025% of |R|)
• R
≥
0
• R
<
0
Type N
• 0.0068 °C + (0.00036% of R)
• All
Summary of Contents for Rosemount 848T
Page 2: ......
Page 4: ...ii Reference Manual 00809 0100 4697 Rev GA Title Page November 2014 ...
Page 12: ...4 Reference Manual 00809 0100 4697 Rev GA Section 1 Introduction November 2014 Introduction ...
Page 26: ...18 Reference Manual 00809 0100 4697 Rev GA Section 2 Installation November 2014 Installation ...
Page 134: ...Index 4 Reference Manual 00809 0100 4697 Rev GA Index November 2014 Index ...
Page 135: ......