370XA Gas Chromatograph
System Reference Manual
APRIL 2014
7P00370-H01
4 - 29
4.3.5.4
The Calibration screen
You must be logged in at least at the Regular user level to access this
screen. If you are not already logged in, the Login screen will display.
There are two types of calibration:
•
Select
Normal
to perform a manual calibration in which the
component data table for the selected stream will be updated
with calibration data unless the data is outside the acceptable
deviations that are listed in the component data table.
•
Select
Forced
to perform a manual calibration in which the
component data table for the selected stream will be updated
with calibration data even if that data is outside the acceptable
deviations that are listed in the component data table.
Select the
Purge stream for 60 seconds
check box if you want to allow
sample gas to flow through the sample loop for 60 seconds prior to
beginning the first analysis.
To start a calibration, select a stream and then press
Enter
on the
keypad.
4.3.5.5
The Validation screen
You must be logged in at least at the Regular user level to access this
screen. If you are not already logged in, the Login screen will display.
Select the
Purge stream for 60 seconds
check box if you want to allow
sample gas to flow through the sample loop for 60 seconds prior to
beginning the first analysis.
To start a validation, select a stream and then press
Enter
on the
keypad.
Summary of Contents for Rosemount 370XA
Page 1: ...Revision A 7P00370 H01 April 2014 370XA Gas Chromatograph System Reference Manual ...
Page 32: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 5 3 3 2 2 Dimensions ...
Page 137: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 5 8 ...
Page 138: ...6 1 Section 6 Overhauling the maintainable module ...
Page 177: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 40 Column 2 ...
Page 178: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 41 Column3 ...
Page 179: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 42 Column 4 ...
Page 182: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 45 Sample Loop SL ...
Page 195: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 8 ...
Page 198: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 11 ...
Page 199: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 12 ...
Page 200: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 13 ...
Page 202: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 15 ...
Page 204: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 17 ...
Page 205: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 18 ...
Page 206: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 19 ...