370XA Gas Chromatograph
System Reference Manual
APRIL 2014
7P00370-H01
3 - 25
3.6
Actuation gas
The analytical valves require actuation gas to operate. When helium
is used as a carrier gas, the default configuration is to also use helium
as the actuation gas. When hydrogen is used as a carrier gas, or if
there is a desire to minimize helium usage, another clean dry gas
such as nitrogen or clean-dry air should be used as the actuation gas.
NOTICE
If you intend to use locally generated instrument air, ensure that the pressure is
sufficient and use filters and dryers to ensure the actuation gas will meet the above
specifications in order to avoid excessive maintenance.
3.6.1
Helium actuation gas
When the carrier gas is also used as the actuation gas, the actuation
gas supply connection should be teed from the helium supply after the
moisture filter.
3.6.2
Alternative actuation gas
If a gas other than the carrier gas is to be used as the actuation gas,
the supply should be connected directly to the actuation gas port on
the 370XA gas manifold. The actuation gas should be nitrogen, dry
air, or some other non-hazardous gas.
Actuation gas specifications
Moisture content
Less than 5 ppm
Particulate
Less than 2 microns
Supply pressure
90 PSI (620 kPaG)
Summary of Contents for Rosemount 370XA
Page 1: ...Revision A 7P00370 H01 April 2014 370XA Gas Chromatograph System Reference Manual ...
Page 32: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 5 3 3 2 2 Dimensions ...
Page 137: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 5 8 ...
Page 138: ...6 1 Section 6 Overhauling the maintainable module ...
Page 177: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 40 Column 2 ...
Page 178: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 41 Column3 ...
Page 179: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 42 Column 4 ...
Page 182: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 45 Sample Loop SL ...
Page 195: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 8 ...
Page 198: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 11 ...
Page 199: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 12 ...
Page 200: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 13 ...
Page 202: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 15 ...
Page 204: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 17 ...
Page 205: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 18 ...
Page 206: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 19 ...