370XA Gas Chromatograph
System Reference Manual
APRIL 2014
7P00370-H01
3 - 57
12.
Make sure the
Purge Stream for 60 seconds
check box and
the
Normal
check box are selected.
13.
Press
to start the calibration.
14.
Wait for the calibration cycle to complete. By default, this will
run for 3 analysis cycles for a total of 13 minutes, including the
60-second purge.
NOTICE
If the calibration generated any alarms go to the
Current Alarms
screen to view
them. Refer to the troubleshooting section to learn how to resolve the calibration
issues.
15.
Open the isolation valves for the sample stream(s) and set the
pressure to between 15 and 30 PSI (100 to 200 kPa).
16.
Go to the
GC Control
menu and select
Auto Sequence
. The
Start Auto Sequence
screen displays.
17.
Make sure the
Purge Stream for 60 seconds
check box is
selected.
18.
Press
to start the analysis.
19.
Go to the
View
menu and select
Reports
. The
Report Display
screen displays.
20.
View the
Analysis Report
for each stream and confirm that
the un-normalized total is between 98 and 102. If it is not, refer
to the troubleshooting section.
The gas chromatograph is now running and analyzing the sample
streams. It will automatically calibrate once a day with the default
settings, so the calibration gas must remain on. Refer to the
Advanced Configuration and Operation section for further information
and configuration instructions.
Summary of Contents for Rosemount 370XA
Page 1: ...Revision A 7P00370 H01 April 2014 370XA Gas Chromatograph System Reference Manual ...
Page 32: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 5 3 3 2 2 Dimensions ...
Page 137: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 5 8 ...
Page 138: ...6 1 Section 6 Overhauling the maintainable module ...
Page 177: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 40 Column 2 ...
Page 178: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 41 Column3 ...
Page 179: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 42 Column 4 ...
Page 182: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 45 Sample Loop SL ...
Page 195: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 8 ...
Page 198: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 11 ...
Page 199: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 12 ...
Page 200: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 13 ...
Page 202: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 15 ...
Page 204: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 17 ...
Page 205: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 18 ...
Page 206: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 19 ...