370XA Gas Chromatograph
System Reference Manual
APRIL 2014
7P00370-H01
7 - 5
7.2
Flowcharts
If you do not find help here contact your local Rosemount Analytical
service representative for help.
Stream 1 RF Out Of Order
Streams 1 to 3 should only be set to Analysis or Not Used.
This alarm may occur If the streams are incorrectly
configured for calibration. To resolve this error, start
MON2020 and select
Streams
on the
Application
menu.
Set the stream’s
Usage
to
Analysis
or
Unused
.
Stream 2 RF Out Of Order
Stream 3 RF Out Of Order
Stream 4 RF Out Of Order
The response factors for one or more components is not
in the order of thermal conductivity configured for natural
gas applications.
Calibration Energy Value Check Fail
The energy value calculated from the analysis value of
the calibration during a calibration cycle is not within the
limits of the energy value entered from the calibration
certificate.
Stored Data Integrity Failure
Data associated with the GC’s measurement results and
logs are stored in the GC along with a 16-bit cyclic
redundancy check (CRC) code. When the data is
retrieved, its integrity is verified by recomputing the CRC
code and comparing it with the CRC code stored along
with the data. If there is a mismatch, then this alarm is
generated, and you should save the current configuration
and replace the CPU board.
n-Pentane RT Drift
The retention time for the n-Pentane peak has drifted too
close to valve 3’s timing. This is typically due to
contamination in the valves and indicates the module
should be exchanged and overhauled soon.
Ethane RT Drift
The retention time for the ethane peak has drifted too
close to the end of the analysis cycle. This is typically due
to contamination in the valves and indicates the module
should be exchanged and overhauled soon.
Alarm
Description and Suggested Actions
Summary of Contents for Rosemount 370XA
Page 1: ...Revision A 7P00370 H01 April 2014 370XA Gas Chromatograph System Reference Manual ...
Page 32: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 5 3 3 2 2 Dimensions ...
Page 137: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 5 8 ...
Page 138: ...6 1 Section 6 Overhauling the maintainable module ...
Page 177: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 40 Column 2 ...
Page 178: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 41 Column3 ...
Page 179: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 42 Column 4 ...
Page 182: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 45 Sample Loop SL ...
Page 195: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 8 ...
Page 198: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 11 ...
Page 199: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 12 ...
Page 200: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 13 ...
Page 202: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 15 ...
Page 204: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 17 ...
Page 205: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 18 ...
Page 206: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 19 ...