370XA Gas Chromatograph
System Reference Manual
APRIL 2014
7P00370-H01
3 - 53
3.14.9
Configure the serial communications settings from
the 370XA
To configure the serial port communication settings to communicate
with ModBus host devices such as a flow computer, RTU, SCADA, or
DCS, the protocol settings for all the devices on the network must
match. Obtain the required serial port settings prior to configuring the
settings on the GC.
1.
Go to the Main Menu and select
Communications
on the
Application
menu. The
Communications
screen opens.
2.
Enter the serial port settings.
(a.)
Modbus ID
- The address that the host device will use
to communicate with the GC. For applications where the
GC is the only slave device on the network, the ModBus
ID is typically set to 1. For multi-dropped applications
where the GC is one of several on the serial network,
the ModBus ID needs to be unique. Refer to your host
device configuration to determine the ModBus ID to be
configured on the GC.
(b.)
Baud Rate
- The baud rate can be set at the standard
rates from 1200 baud up to 57600. For ModBus commu-
nications, the typical setting is
9600
.
(c.)
Data Bits
- The number of bits used for communica-
tions. The typical setting for ASCII mode
communications is
7
. The typical setting for RTU mode
communications is
8
.
(d.)
Stop Bits
- The number of bits sent to indicate the end
of a message. Typically set to
1
.
(e.)
Parity
- The error checking mode for the parity bit in
ASCII mode messages. This can be set to either
ODD
or
EVEN
for ASCII mode communications and must
match the host device’s setting. Set to
NONE
for RTU
mode communications.
(f.)
MAP File
- The ModBus address map. By default, this is
set to SIM_2251, which is a pre-configured map with the
same mapping as the Daniel 2251 controller, and is the
most common communication mapping for flow
Summary of Contents for Rosemount 370XA
Page 1: ...Revision A 7P00370 H01 April 2014 370XA Gas Chromatograph System Reference Manual ...
Page 32: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 5 3 3 2 2 Dimensions ...
Page 137: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 5 8 ...
Page 138: ...6 1 Section 6 Overhauling the maintainable module ...
Page 177: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 40 Column 2 ...
Page 178: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 41 Column3 ...
Page 179: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 42 Column 4 ...
Page 182: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 45 Sample Loop SL ...
Page 195: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 8 ...
Page 198: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 11 ...
Page 199: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 12 ...
Page 200: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 13 ...
Page 202: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 15 ...
Page 204: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 17 ...
Page 205: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 18 ...
Page 206: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 19 ...