The following image shows the carrier gas connector (1).
The carrier gas must be regulated from bottle pressure to 90 PSI (620
kPa) using a two-stage bottle regulator with stainless steel
diaphragms. Use a dual-stage regulator to ensure the outlet pressure
will not change with changes in the bottle pressure. Stainless steel
diaphragms are required to avoid contamination of the analytical
oven. Use
⅛
-inch stainless steel tubing that is clean and free of
grease to connect from the carrier gas bottle manifold to the inlet of
the carrier gas drier (1) on the 370XA sample conditioning system (as
shown [show single stream and multi stream panels and indicate inlet
of drier). Before making the final connection to the sample system,
blow through the lines with helium for 30 seconds to remove any
contamination such as water or metal shavings from cutting the tube.
WARNING
Regulate the carrier gas to 90 PSI (620 kPa). Pressures higher than this may
damage the analyzer and cause an unsafe environment.
Summary of Contents for Rosemount 370XA
Page 1: ...Revision A 7P00370 H01 April 2014 370XA Gas Chromatograph System Reference Manual ...
Page 32: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 5 3 3 2 2 Dimensions ...
Page 137: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 5 8 ...
Page 138: ...6 1 Section 6 Overhauling the maintainable module ...
Page 177: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 40 Column 2 ...
Page 178: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 41 Column3 ...
Page 179: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 42 Column 4 ...
Page 182: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 45 Sample Loop SL ...
Page 195: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 8 ...
Page 198: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 11 ...
Page 199: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 12 ...
Page 200: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 13 ...
Page 202: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 15 ...
Page 204: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 17 ...
Page 205: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 18 ...
Page 206: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 19 ...