System Reference Manual
370XA Gas Chromatograph
7P00370-H01
APRIL 2014
3 - 50
3.14.5
Set the time
1.
Press the enter key to go to the Main Menu.
2.
Use the right arrow key to move to the Tools menu.
3.
Use the down arrow key to move to the Set GC Time command
and press the enter key. The Set GC Time screen displays.
4.
Set the current date and time.
•
Press the edit key to activate a field.
•
Press the enter key to accept an entry and to de-activate
the field.
•
Use the arrow keys to move to the next field.
5.
If your country does not employ daylight savings time, clear the
Enable Day Light Savings
check box, which is checked by
default.
NOTICE
You must use MON2020 to configure daylight savings time.
3.14.6
Configure the gas chromatograph’s Ethernet ports
1.
Go to the Main Menu and select
TCP/IP
on the
Application
menu. The
TCP/IP
screen displays.
2.
Make a note of the Ethernet settings for both ports.
Ethernet 1
is the RJ-45 terminal that is commonly used for local computer
access;
Ethernet 2
is the port that is commonly used for
communication with a supervisory system such as a flow
computer, RTU, SCADA, or DCS.
3.
Enter the Ethernet settings according to the network
requirements of your installation. If you intend to use
Ethernet
1
for local access only, do not change the settings. Contact
Summary of Contents for Rosemount 370XA
Page 1: ...Revision A 7P00370 H01 April 2014 370XA Gas Chromatograph System Reference Manual ...
Page 32: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 5 3 3 2 2 Dimensions ...
Page 137: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 5 8 ...
Page 138: ...6 1 Section 6 Overhauling the maintainable module ...
Page 177: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 40 Column 2 ...
Page 178: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 41 Column3 ...
Page 179: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 42 Column 4 ...
Page 182: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 45 Sample Loop SL ...
Page 195: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 8 ...
Page 198: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 11 ...
Page 199: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 12 ...
Page 200: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 13 ...
Page 202: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 15 ...
Page 204: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 17 ...
Page 205: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 18 ...
Page 206: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 19 ...