75
Reference Manual
00809-0100-2460, Rev DE
October 2018
4.3.6
System setup
The System Values window lets you specify parameters and units for inventory calculations.
1. Log in to the Web interface as described in
“Using the web interface” on page 114
.
2. Select
Configuration>System Values
.
Figure 4-7. System Parameters and Units
Manual values
Select the appropriate check boxes in case you like to use manual values for ambient air temperature and
pressure, and type the desired values into the input fields.
Reference temperature
The Rosemount 2460 System Hub performs inventory calculations according to the API Manual of
Petroleum Measurement Standards Chapter 12, Section 1, at the standard reference temperature 15°C
(60°F). This is the default reference temperature.
Other reference temperatures can be specified in the Reference Temperature input field. Ensure that
correct RT volume table, for example 54B-2004, is used for the product.
System units
Level, Level Rate, Temperature, and Pressure units are configured in TankMaster WinSetup configuration
program.
Summary of Contents for Rosemount 2460
Page 2: ......
Page 4: ...ii Reference Manual 00809 0100 2460 Rev DE Title Page October 2018 Title Page ...
Page 12: ...x Reference Manual 00809 0100 2460 Rev DE Contents October 2018 Contents ...
Page 18: ...6 Reference Manual 00809 0100 2460 Rev DE Introduction October 2018 Introduction ...
Page 60: ...48 Reference Manual 00809 0100 2460 Rev DE Installation October 2018 Installation ...
Page 100: ...88 Reference Manual 00809 0100 2460 Rev DE Configuration October 2018 Configuration ...
Page 238: ...226 Reference Manual 00809 0100 2460 Rev DE Index October 2018 Index ...
Page 239: ......