H.2
Advanced L&J configuration
This section describes advanced configuration options for supported L&J devices.
H.2.1
Fast polling
for information on how to setup fast polling.
H.2.2
Advanced L&J configuration
1. Open the TankMaster WinSetup program.
2. In the WinSetup workspace, click the right mouse button on the Rosemount 2460 icon and choose
Properties.
3. Select the Tank Database tab.
4. For the desired tank position with L&J device, press the
Advanced
button to open the
L&J Configuration window:
5. The L&J Configuration window lets you configure advanced options for the selected device.
6. Press
OK
to save the current configuration and close the configuration window.
Option
Description
Inverse Temperature Sign
This option shifts sign for Average Liquid Temperature values.
Use this option to match host system standards.
Swap Interface and Pressure
This option shifts interface and pressure data.
Use this option to match host system standards.
No Response If Invalid Level
This option prevents invalid level data to result in communication failure
(CFail) error message.
Advanced button
L&J device
Summary of Contents for Rosemount 2460
Page 2: ......
Page 4: ...ii Reference Manual 00809 0100 2460 Rev DE Title Page October 2018 Title Page ...
Page 12: ...x Reference Manual 00809 0100 2460 Rev DE Contents October 2018 Contents ...
Page 18: ...6 Reference Manual 00809 0100 2460 Rev DE Introduction October 2018 Introduction ...
Page 60: ...48 Reference Manual 00809 0100 2460 Rev DE Installation October 2018 Installation ...
Page 100: ...88 Reference Manual 00809 0100 2460 Rev DE Configuration October 2018 Configuration ...
Page 238: ...226 Reference Manual 00809 0100 2460 Rev DE Index October 2018 Index ...
Page 239: ......