Hardness Tester
Startup and Retooling with
Optional Accessories
DuraJet 10
7-5
General safety instructions:
DANGER
Electric shock
Prior to starting up the machine, you must make a safety compliance assessment in
accordance with Machinery Directive 2006/42/EC. Comply with all the labour safety and
accident prevention rules and regulations that are applicable at the location where the
machine is operated. The external machine control system (EMERGENCY STOP and Z-
axis downward or clamping) must satisfy at least PL [d] to EN 13849-1. Safely de-energize
the entire system before you install, disassemble, or wire the external machine control
system. Only qualified technical personnel are allowed to install, assemble, start up,
maintain, and shut down safety components.
Startup:
Validation to DIN EN ISO 13849-2 must be performed. Startup must be documented.
There must be proof of the safety functions.