Hardness Tester
Advanced Options
DuraJet 10
5-11
Remove the nose cone.
Remove the indenter. In its position, install the special indenter that is usually
delivered together with the depth meter (for turret option, install at position 1).
Select the
Depth probe
menu.
In the
Select preload
box, select the preload for which you wish to calibrate.
Place the measuring device of the depth meter on an appropriate support under the
test unit. During the calibration of the depth probe, the tip of the indenter touches
the measuring device of the depth meter.
Move either the measuring device or the test unit until the indenter is located just
above the measuring device (approximately 0.5 mm).
Select the
Start
menu.
Depth probe menu
After the set load has been reached, a text message appears below the progress bar
prompting you to perform a test.
Touch the
Reset
button to set the value of the depth probe to zero and start your
calibration as specified in the standard.
As soon as you have completed your calibration, press the
Exit
button, and the test
of the depth probe is terminated.
5.10 Documenting and Evaluating Results
5.10.1 Displaying, Printing and Exporting Results
After you have carried out the measurements, you can view them in the archive or
directly during the test in the List tab.
Each list of measured values can accommodate a maximum of 500 measured values.
The oldest measured value is overwritten automatically when the list is full.
Results