Hardness Tester
Advanced Options
DuraJet 10
5-10
Load calibration menu
In the boxes
Test procedure
and
Test method
, select the test procedure and the
test method for which you want to check load calibration.
Place the load cell of the calibration equipment under the test unit on the test anvil
of your support.
In load calibration, the tip of the indenter touches the load cell of the calibration
equipment. For this reason, it is advisable to place a workpiece on the load cell as an
intermediate layer.
NOTE
Risk of damage
Always perform calibration in unclamped state; otherwise there is already a force that
acts on the load cell.
Place a workpiece on the load cell as an intermediate layer, and set the display of
the calibration equipment to zero.
Press the
nose cone
button on the control unit to move the nose cone until it
stands just above the load cell with the workpiece (do not clamp!).
Select the
Start
button.
Read the measured value on the calibration equipment, and enter the difference as
correction value in the boxes provided.
5.9
Calibrating the Depth Gauge System
You need an external depth meter to calibrate the depth probe. To calibrate the depth
probe, the machine must always be unclamped and without nose cone.