Hardness Tester
Startup and Retooling with
Optional Accessories
DuraJet 10
7-2
The USB flash disk (USB stick) may have only a single partition.
7.2
External machine control via hardware interface
For this option the DuraJet G5 is equipped with an additional interface at the rear of the
machine stand.
Via this interface it is possible to control the DuraJet G5 via signals for "Start
measurement", "Clamp test unit", "Unclamp test unit" and "Emergency stop".
Likewise, the outputs "Classification OK", "Classification High", "Classification Low",
"Cycle active" and "Nose cone clamped" are provided.
The intended field of application for this option is, for example, the integration of a
DuraJet G5 into a production plant.
If this option is ordered, the shipment additionally includes 1 activation plug and 1 blank
plug.
The activation plug is needed to use the machine as a stand-alone unit. This means that
the hardness tester works only when the activation plug is inserted, or when the
machine is, with the blank plug provided, integrated in an external controller with a
proper safety structure. If a plug is not connected to the interface, the hardness tester is
not functional, since the safety circuit is interrupted.
Only the customer is allowed to integrate the unit with a blank plug into a customer
system.