Hardness Tester
Startup and Retooling with
Optional Accessories
DuraJet 10
7-3
Activation plug or blank plug
Signals - interface:
Inputs:
▪
Measurement start
▪
Clamping test unit
▪
Unclamping test unit
▪
Emergency OFF
Outputs:
▪
Classification OK
▪
Classification high
▪
Classification low
▪
Cycle active
▪
Nose cone is clamped
Scope of supply:
In addition the the standard scope of supply of a DuraJet G5, the following parts are
enclosed:
▪
1x activation plug
▪
1x blank plug
Instructions for wiring and cable routing:
The +24VDC power supply line (pin 1) is protected with a 3-A fuse that is installed in the
DuraJet G5 unit. The safety lines should be routed such that they are protected and
separate from the other lines. Perform cable routing, cable selection and dimensioning
in compliance with the regulations applicable on site. All interconnections must be via
floating connections.