Common notes
Target system
is the physical Printed Circuit Board (PCB), which contains the device to be
in-system programmed.
ISP programmer
is programmer, which has in-system programming capability (for example
BeeProg2, BeeProg2C, , SmartProg2 …).
General rules for in-system programming
We recommended respect following rules to avoid damage PC, ISP programmer, and target
device or target system:
•
Ensure common earth point for target system, ISP programmer and PC.
•
For laptop or other PC that is not connected to common earth point: make hard - wired
connection from laptop to common earth point (for example use LPT or COM port D –
connector).
•
Any devices connected to target system must be connected to common earth point too.
Direction of connect ELNEC ISP programmer to target system:
During in-system programming you connect two electrical devices – ISP programmer and
target system. Unqualified connection can damage these devices.
Note:
When you don’t keep below directions and you damage programmer during in-system
programming, it is damage of programmer by unqualified manipulation and is out of warranty.
1. Turn off both devices – ISP programmer and target device.
2. Assign same GND potential for all devices, e.g. connect GND of all devices by wire.
3. Insert one connector of ISP cable to ISP programmer, turn on programmer and control
program.
4. In control program select target device and operation options.
5. Start action on target device (read, program).
6. After direction of control program, connect other ISP cable connector to target system and
turn on it.
7. After direction of control program, disconnect other ISP cable connector from target system
and turn off it.
8. If you need another action on target device, you continue with step 5.
The recommendation for design of target system with ISP
programmed device
The target system must be designed to allow all signals, which are use for In-system
programming to be directly connected to ISP programmer via ISP connector. If target system
use these signals for other function, is necessary isolated these signals. Target system
mustn’t affect these signals during In-system programming.
For in-system programmable devices manufacturers publish application notes. Design of
ELNEC programmers together with respect of these application notes allows proper In-
system programming. Condition is exactly respecting these application notes. Applications
163
Summary of Contents for SEEprog
Page 7: ...Introduction Introduction 7...
Page 12: ...ELNEC s r o Quick Start 12...
Page 15: ...Detailed description Detailed description 15...
Page 16: ...ELNEC s r o BeeHive204 16...
Page 30: ...ELNEC s r o BeeProg2 BeeProg2C BeeProg 30...
Page 45: ...SmartProg2 SmartProg2 45...
Page 54: ...ELNEC s r o MEMprog2 54...
Page 61: ...SEEprog SEEprog 61...
Page 66: ...ELNEC s r o Setup 66...
Page 77: ...PG4UW PG4UW 77...
Page 144: ...ELNEC s r o PG4UWMC 144...
Page 154: ...ELNEC s r o Common notes 154...
Page 168: ...ELNEC s r o Troubleshooting and warranty 168...