PG4UW
Pins of programmer's ZIF excluded from sensing
contains list of pins that will be ignored
from testing by Automatic YES!. The reason to ignore the pins is mostly - capacitors
connected to these pins.
Button
Setting Automatic YES! parameters
will run wizard that can detect permanently
connected pins (pins with capacitors) and set these pins to list of pins excluded from sensing.
After selecting of device, list of excluded pins contains default excluded pins for selected
device adapter. If other bypass capacitors to universal programmer and/or device adapter are
added by customer, there is necessary to run
Automatic YES! parameters wizard
to
override default parameters and detect other pins with capacitors.
In
Device removal hold off time
is time period between you removed device from the ZIF
socket and the time when software starts to check the socket for new device inserted. This
interval is in seconds and must be from 1 to 120 (default value is 2 seconds).
In
Device insertion complete time
is possible to set a time within all pins of the device have
to be properly inserted after a first pin(s) detected so that the program will not detects
incorrectly inserted device. This interval is in seconds and must be from 1 to 120 (default
value is 5 seconds).
The
Suspend on error
defines if the Automatic YES! function will be temporary disabled on
error to see result of operation or will going on without suspension.
The options are set to defaults after new device is selected by
Device / Select device
This setting is saved to disk by command
Options / Save options
and could be saved into
the project file for selected device File / Save project ...
Note:
When using device socket adapters with some passive or active parts, for example
capacitors for bypassing supply voltage, the Automatic YES! function may need to set these
pins to Pins with capacitors list. This is necessary to make Automatic YES! function working
properly. Otherwise Automatic YES! function will "think" the pins are still connected and it will
not allow user to insert new device and start new programming.
Programmer / Selftest
Command executes a selftest of current programmer without diagnostic POD. We strongly
recommend execute also
Programmer / Selftest
plus
of programmer, because Selftest
procedure without diagnostic POD is not able to check whole programmer and to discover (if
exist) some special errors.
Programmer / Selftest plus
Command executes a selftest plus of current programmer using diagnostic POD, which is
included in standard delivery of programmer.
Recommendations how often run Selftest plus you can find at Maintenance section.
127
Summary of Contents for SEEprog
Page 7: ...Introduction Introduction 7...
Page 12: ...ELNEC s r o Quick Start 12...
Page 15: ...Detailed description Detailed description 15...
Page 16: ...ELNEC s r o BeeHive204 16...
Page 30: ...ELNEC s r o BeeProg2 BeeProg2C BeeProg 30...
Page 45: ...SmartProg2 SmartProg2 45...
Page 54: ...ELNEC s r o MEMprog2 54...
Page 61: ...SEEprog SEEprog 61...
Page 66: ...ELNEC s r o Setup 66...
Page 77: ...PG4UW PG4UW 77...
Page 144: ...ELNEC s r o PG4UWMC 144...
Page 154: ...ELNEC s r o Common notes 154...
Page 168: ...ELNEC s r o Troubleshooting and warranty 168...