Sequence for testing 20 pins ISP connector:
1. Insert Diagnostic POD for ISP connectors #2 into ZIF socket of the programmer.
Diagnostic POD for ISP connectors #2 must be inserted as 48 pins device.
2. Interconnect 20 pins connector of Diagnostic POD for ISP connectors #2 with an ISP
connector of the programmer with an ISP cable, included in delivery programmer
package. Be sure that pins are interconnected properly (i.e. 1-1, 2-2, ..., 20-20).
3. Run selftest of ISP connector in PG4UW (Programmer / Selftest ISP connector…).
We recommend run this test every 6 months.
Programmer / Calibration test
Command executes test of programmer's calibration values using
48 Pins Calibration test
POD, Type I,
which is optional accessories for BeeHive204, BeeProg2, BeeProg2C and
. The order number: 70-0438.
There are tested voltage levels of TTL drivers, VCCP, VPP1 and VPP2 voltages on each pin
of ZIF socket. Result of the Calibration Test can be saved into file and/or printed (for next
use).
Programmer / Common notes
Serial numbering
Table bellow is showing differences between old and new serial numbering of programmers:
new numbering
old numbering
050-00xxx 050-xxx
050-01xxx 950-xxx
050-02xxx 850-xxx
050-03xxx 750-xxx
129
Summary of Contents for SEEprog
Page 7: ...Introduction Introduction 7...
Page 12: ...ELNEC s r o Quick Start 12...
Page 15: ...Detailed description Detailed description 15...
Page 16: ...ELNEC s r o BeeHive204 16...
Page 30: ...ELNEC s r o BeeProg2 BeeProg2C BeeProg 30...
Page 45: ...SmartProg2 SmartProg2 45...
Page 54: ...ELNEC s r o MEMprog2 54...
Page 61: ...SEEprog SEEprog 61...
Page 66: ...ELNEC s r o Setup 66...
Page 77: ...PG4UW PG4UW 77...
Page 144: ...ELNEC s r o PG4UWMC 144...
Page 154: ...ELNEC s r o Common notes 154...
Page 168: ...ELNEC s r o Troubleshooting and warranty 168...