ADR-Auto Range
29
9901X0215 Issue 5
The screen display will now be of the form –
Press the
Up Arrow
key to return to the previous screen.
Use the
Left
and
Right Arrow
keys to step backwards and forwards through the
records. Note that the records are numbered, and that the oldest record is always
number 1. To find the newest records, press the
Right Arrow
key repeatedly. To
find the oldest record, press the
Left Arrow
key.
To return to the
Initial Screen
, press the
Escape
key.
4.10
Verifying the Measurements
To check the ADR-Auto’s calibration, press the
Menu
key to display the
Installation
and Setup
options. First select the operating mode required and then select
Verify
Mode
. The display will be of the form –
Use the
Left
or
Right Arrow
keys to select which load units are to be used during the
verification and then press the
Enter
key.
The screen display will be of the form –
Place a standard load measuring device in the frame, and press the
Run
key on the
keyboard to turn the main pump on. Note that this also turns the dump valve off.
Using the
Left
and
Right Arrow
keys, open the pump control valve sufficiently to
start the ram rising slowly and press the
Reset
key to tare the load when required. If
a larger change in the pump control valve is required, hold down the
Arrow
key.
Note that the rapid approach pump can be operated using the
Pause
key.
Pressing the
Enter
key at any time freezes the load display with the current load so
that the value may be noted and the message changes to –
Press Enter to Release
_
001 View Records
____________________
↑
Height 150.0
Width 150.0
Depth 150.0
Weight in Air 1000
Weight in Water 2000
Sample Density 1000
__________
Mode One Verification
_________
_
Select the Units
_ SI
_________
Mode One Verification
__________
Load Reading +0.1
Control Valve (%) 0.0
Press ENTER to Freeze
Main Off Rapid Off Dump On