ADR-Auto Range
23
9901X0215 Issue 5
Pace Rate
10 kN Frame
0.05 kN/sec
25 kN Frame
0.05 kN/sec
50 kN Frame
0.1 kN/sec
100 kN Frame
0.1 kN/sec
250 kN Frame
2.4 kN/sec
1000 kN Frame
6.8 kN/sec
1500 kN Frame
6.8 kN/sec
2000 kN Frame
6.8 kN/sec
2225 kN Frame
6.8 kN/sec
3000 kN Frame
6.8 kN/sec
4.7
Running a Test
If the
Initial Screen
is not being displayed, press the
Escape
key until the display is
of the form-
Check that the sample type, and control mode are correct. If they are not, return to
the Setup Mode and set them as required. Before running the test, the Sample
Reference, the Pace Rate and the Sample details can be edited via the
Sample
key.
4.7.1 Setting the Pace Rate
If the pace rate is incorrect, press the
Pace
key. If dual pace rates have been
selected, the screen display will be of the form –
otherwise the display will be of the form –
___
AdrAuto 2000 - Mode 1 Compression
____
Sample Reference TEST-123
Sample Type Cube
Pace Rate 6.80
Control Mode Automatic
11/09/03 14:35:34
____
Edit Pace Rates & Transition Load
___
_
Initial Rate
_ 30.00
Transition Load 500.0
Final Rate 6.80
_______
Edit the Current Pace Rate
_______
_
Pace Rate
_ 6.80