ADR-Auto Range
17
9901X0215 Issue 5
4.6
Configuring the ADR-Auto’s Operating Modes
Before running a test, check that the current operating mode is the required one, and
that the mode has been configured correctly. The operating mode is selected and
configured by pressing the
Menu
key, then
Setup Mode
. Note that the
Installation
and Setup
mode can only be accessed when the
Initial Screen
display is active.
The screen display will be of the form –
Use the
Up
and
Down Arrow
keys to move the cursor to the option required, and
then press the
Enter
key. To return to the
Initial Screen
press the
Escape
key.
4.6.1 Selecting the Operating Mode
When two frames are attached, the ADR-Auto 1500, 2000, 2225 & 3000 is fitted with
a solenoid operated changeover valve inside the console, which is used to connect
the hydraulic power pack to the required frame. If only one frame is attached, this
valve is not fitted.
Note: If the Compression Frame has a front gate with safety interlock, this must
be closed before the solenoid operated changeover valve will operate.
The ADR-Auto 250/25 has a manual frame changeover valve fitted externally (Refer
to figure 1)
To change from one frame to the other, move the flashing cursor to the
Change
Mode
option and press
Enter
. The screen display will be of the form -
Do not attempt to operate the changeover valve until the ram on the currently
selected frame has returned to its initial position. To operate the changeover valve if
the ram is at its initial position, use the
Left
or
Right Arrow
key to select
Yes
and
press the
Enter
key. To exit from this screen without operating the valve either select
No
using the
Left
or
Right Arrow
key and press the
Enter
key, or press the
Escape
key. If a manual changeover valve is fitted, it must be operated manually.
Note that the options set under the
Select Sample
and
Setup Mode
menus apply to
the currently selected operating mode.
Change to the required mode before
setting the options.
4.6.2 Defining the Operating Mode Options
There are several options which affect the operation of the ADR-Auto. To set these,
use the
Up
and
Down Arrow
keys to select the
Setup Mode
option and press
Enter
.
_
Installation and Setup – Select Option
_
_
Select Sample
_ Set Clock
Change Mode Calibrate and Test
Setup Mode Serial Parameters
Reset Mode Display Version No.
Verify Mode
________________
WARNING!
________________
The Rams must be at their Initial
Positions before the Changeover
Valve can Operate. Check them now.
The Current Mode is Mode 1
Change Mode? No