SCD 19102 / SMD 19102
Instruction Manual, 05/2008, A5E01588521A-002
73
Service and maintenance
9
9.1
Cleaning
CAUTION
Device maintenance, cleaning and disinfecting
•
The front panel is extremely sensitive to mechanical damage. Avoid all scratches,
knocks etc.!
SCD / SMD 19102 CP versions with protective glass are considerably more susceptible
to this kind of damage.
•
SCD / SMD 19102 versions, without protective glass: Remove water drops immediately;
extended contact with water discolors the surface.
•
Clean the front panel when dirty using a microfiber cloth and, if necessary, a cleaning
agent. Only clean housing parts using the recommended cleaning agents.
•
The entire display must only be disinfected using the tested disinfectants.
CAUTION
Remove all residual cleaning agents immediately
•
Some cleaning agents leave residue (streaks) on the surface which can easily be
removed by wiping with a damp cloth.
•
Remove residues of alcohol, "Cidex", "Taski DS 5001" and "Surfanios Fraicheur Citron"
from the surface immediately. Residue of these cleaning agents on the surface often
causes permanent streaks.
Recommended cleaning agents and disinfectants:
Agent class
Tested cleaning agents and
disinfectants:
Further examples
Alcohol
•
Ethyl alcohol (96%)
•
Isopropyl alcohol (70%)
•
0.5% chlorhexidine in
isopropyl alcohol (70%)
•
Hospiset cloth
•
Mikrozid liquid
Aldehyde
•
Melsitt
•
Cidex
•
Cidex (2.4% glutaraldehyde
solution)
•
Aldasan 2000
•
Kohsolin
•
Gigasept FF
Summary of Contents for SCD 19102
Page 2: ......
Page 8: ...Table of contents SCD 19102 SMD 19102 8 Instruction Manual 05 2008 A5E01588521A 002...
Page 14: ...Safety notes SCD 19102 SMD 19102 14 Instruction Manual 05 2008 A5E01588521A 002...
Page 76: ...Troubleshooting SCD 19102 SMD 19102 76 Instruction Manual 05 2008 A5E01588521A 002...
Page 103: ......