Commissioning
7.4 Make settings
SCD 19102 / SMD 19102
68
Instruction Manual, 05/2008, A5E01588521A-002
3.
Connect Pixel On / Pixel Off test pattern (e.g. "LCD Adjust" from SMfit® Total Care). If
this is not available, the Pixel On / Pixel Off fields of the SMPTE image can also be used.
4.
Optimize the signal frequency in the "Total horizontal lines" input field so that the vertical
lines of the test pattern are displayed perfectly. It must be possible to recognize
alternating white and black stripes, each with a width of one pixel (see figure).
Figure 7-9
Force Mode example images
5.
Execute "Force"
command using key 2.
Optimize geometry/resolution
Note
Active and inactive resolution range
The active resolution range is defined by the values of the "Horizontal resolution" and
"Vertical resolution" input fields. This range is displayed in black and centric on the display.
The inactive resolution range is the unused range between the active resolution and the
maximum display resolution of 1280 x 1024. This range is displayed in Force Mode in gray
(with monochrome displays) or blue (with color displays).
$FWLYHUHVROXWLRQUDQJH
,QDFWLYHUHVROXWLRQUDQJH
6037(WHVWSDWWHUQ