OPERATION
4.5
Powering ON
Turn "ON" the "MAIN POWER" switch on the power unit. When the power unit is not connected to the
control unit with the DC cable, the breaker will trip. If no error is found, the magnetic bearing functions
and the rotor levitates (POWER ON state). The "POWER" LED illuminates in green (Levitation state).
4.6
Pump operation
4.6.1 Input operation port setting
Set the hardware which operates the STP pump before the operation. The STP pump is equipped with
the parallel port (X2 REMOTE connector), serial port COM1 (X3 COM1 connector), and serial port
COM2 (X4 COM2 connector) as standard hardware for the operation. The STP-Link (optional
accessory) or the display unit iDT-001 (optional accessory) can operate the STP pump via the serial
port COM3 (X5 STP-Link connector), serial port COM4 (X14 STP-Link connector) of the power unit.
When setting the input operation port to the parallel port, refer to Section 4.9, "Parallel port input/output
signal". When setting to a serial port, refer to Section 5, "Serial Communication Protocol" for
information and methods.
The method of operation with the parallel port (X2 REMOTE connector) is shown below. Following it
after reading through Section 4.9, "Parallel port input/output signal". When operating the STP pump via
serial communication, refer to Section 5, "Serial Communication protocol".
When operating the STP pump via the STP-Link (optional accessory) or the display unit iDT-001
(optional accessory), refer to each Instruction Manual.