![Edwards STP Series Instruction Manual Download Page 133](http://html.mh-extra.com/html/edwards/stp-series/stp-series_instruction-manual_2366164133.webp)
STP-iXA2205 Series Turbomolecular Pump
Nov. 10
123
Issue 1-h
5
SERIAL
COMMUNICATION
PROTOCOL
Example:
Pump operation mode
: 01
hex
= 1 = Levitation
WARNING being detected : 000C
hex
= 0004
hex
OR 0080
hex
=
"WARNING: First Damage Limit" and
"WARNING: Imbalance X_H"
The number of error
: 02
hex
= 2 errors
Error 1
: 0D
hex
= 13 = Disturbance Xh
Error 2
: 0F
hex
= 15 = Disturbance Xb
Error 3 to 32
: 00
hex
= No error detected
Parameter
1
2
3
4
5
6
7
8
9
10
ASCII
0 1 0
0 0 C 0
2
0 D 0 F 0
0
0
0
0
0 0 0 0 0
HEX
30 31 30 30 30 43 30 32 30 44 30 46 30 30 30 30 30 30 30 30 30 30
Parameter
11
12
13
14
15
70
71
72
73
74
ASCII
0 0 0
0 0 0 0
0
0
0
0
0
0
0
0
0 0 0 0 0
HEX
30 30 30 30 30 30 30 30 30 30
[Omitted]
30 30 30 30 30 30 30 30 30 30
Parameter
75
76
77
78
79
80
81
82
ASCII
0 0 0
0 0 0 0
0
0
0
0
0
0
0
0
0
HEX
30 30 30 30 30 30 30 30 30 30 30 30 30 30 30 30