STP-iXA2205 Series Turbomolecular Pump
Nov. 10
62
Issue 1-h
INSTALLATION
3
3.13.6 Connecting the pump and TMS valve
Connect the pump and TMS valve according to Figure 46.
(A) Two-way valve type
(B) Three-way valve type
Figure 46 - TMS valve connection
Use cooling water which fulfils the conditions in Section 2.3, “Water cooling use condition”.
Note: When the two-way vale is closed, cooling water circulating whole device will stop.
Note: Procure and connect the cooling water pipe and affix the electromagnetic cooling water valve
at your site. The diameter of the valve is Rc1/4.
3.13.7 TMS unit specification
Item
Condition
Temperature control method
Control ON/OFF of the TMS heater and cooling water
Setting temperature
°C
Standard type: 70
Failure output
Failure outputs from the STP control unit
TMS heater voltage
ACV (Hz)
220 (50/60)
TMS heater capacity
W
300
TMS valve voltage
DCV
24
Table 13 - TMS unit specification
Equipment
STP pump
STP pump
Equipment