Remedy
1. Examine the equipment recently connected to the pump. Make sure that is
installed and correctly connected.
2. If the warning does not disappear power cycle the pump.
3. If the warning re-appears after power cycling, contact us.
Fault
Event 258.63 - USB V oload
Cause
Faulty equipment has just been plugged in
Remedy
1. Examine the equipment recently connected to the pump. Make sure that is
installed and correctly connected.
2. If the warning does not disappear power cycle the pump.
3. If the warning re-appears after power cycling, contact us.
8.5 Inverter warnings and alarms
The inverters used to drive the dry pump and booster can also give warnings and alarms
if problems are identified. Inverter warning and alarm codes are displayed by the PDT if
installed.
Table 38
Inverter warning and alarm status registers
Inverter warning
and alarm status
registers
Text displayed on
PDT as alert and fault
history
Description
Recommended action
0123456789012
0800 0000
NO_INV_PARAMS
No inverter parameters set loa-
ded.
No inverter parameters
set loaded.
0000 0010
POWER_DOWN___
Supply voltage collapsed (power
off)
Check Supply Voltage
0000 0008
POWER_DOWN___
Supply voltage collapsed (power
off)
Check Supply Voltage
0040 0000
EMS_ACTIVATED
EMS activated
EMS activated
0080 0000
MISSING_PHASE
Missing Phase in mains supply
Missing Phase in mains
supply
0000 0200
MTR_TEMP_HIGH
High Motor Temperature
Check cooling
0000 0020
MTR_TEMP_HIGH
High Motor Temperature
Check cooling
0000 0400
INV_TEMP_HIGH
High Inverter Sink Temperature
Check cooling
0000 0040
INV_TEMP_HIGH
High Inverter Sink Temperature
Check cooling
0000 0800
INV_TEMP_HIGH
High Inverter Processor Temper-
ature
Check cooling
0008 0000
INV_TEMP_HIGH
High Inverter Temperature Trip
Check cooling
0000 0100
MISSING_PHASE
Missing Phase in mains supply
Missing Phase in mains
supply
8000 0000
LOW_SPEED____
Acceleration timeout (Pump
Seized/overload)
Acceleration timeout
(Pump Seized/overload)
10/2020 - ©Edwards Limited
Page 83
M56635880_D
M56635880_D - Fault finding
Summary of Contents for iXM Series
Page 1: ...Dry Pump System iXM INSTRUCTION MANUAL M56635880_D Original instructions edwardsvacuum com...
Page 101: ...This page has been intentionally left blank...
Page 102: ...This page has been intentionally left blank...
Page 103: ...This page has been intentionally left blank...
Page 104: ...edwardsvacuum com...