3.9 Tracer gas analysis
Tracer gas fugitive emission testing has been carried out in accordance with the SEMI S6
Standard.
Table 13
Tracer gas test parameters
Test parameters
Tracer gas
N
2
O (Nitrous Oxide)
Tracer gas concentration
100%
Tracer gas release rate
5 slpm (standard litres per minute)
Tracer gas release points
Exhaust flange on the iXM200/iXM600/iXM1200 pump
Process gas
Nitrogen
Table 14
Tracer gas test system parameters
System parameters
Units
iXM3000 iXM1200
iXM600
iXM200
Extraction flow rate:
From port on top of enclosure
From port on exhaust extraction kit
m
3
/h
396.4
107.7
121.2
Volume of enclosure
m
3
0.423
0.122
0.077
Free air volume of enclosure
m
3
0.265
0.072
0.040
Air changes per minute
24.9
27.9
50.0
Hardware configuration:
▪ 100 mm duct connected to port on
top of enclosure
▪ Exhaust extraction cover kit
installed
Yes
Yes
Yes
Yes
Yes
Yes
Note:
The recommended extraction flow rate for iXM200, iXM600, iXM1200 is 180 m
3
/h.
Table 15
Worst case test results
Process gas
Chemical
name
Maximum
process
gas flow
(slm)
TLV/LEL
(ppm)*
25%
TLV/LEL
(ppm)
N
2
O re-
lease
rate
(slm)
Maximum
N
2
O de-
tected
outside
enclosure
(ppm)
ERC
(ppm) Pass/Fail†
Difluoromethane CH
2
F
2
0.2
1000
250
5
3
0.12
Pass
Methane
CH
4
1
1000
250
5
3
0.60
Pass
Carbon Monox-
ide
CO
0.5
25
6.25
5
3
0.30
Pass
Hydrogen
H
2
40
4000
1000
5
3
24.00
Pass
Nitrous Oxide
N
2
O
15
50
12.5
5
3
9.00
Pass
10/2020 - ©Edwards Limited
Page 23
M56635880_D
M56635880_D - Technical data
Summary of Contents for iXM Series
Page 1: ...Dry Pump System iXM INSTRUCTION MANUAL M56635880_D Original instructions edwardsvacuum com...
Page 101: ...This page has been intentionally left blank...
Page 102: ...This page has been intentionally left blank...
Page 103: ...This page has been intentionally left blank...
Page 104: ...edwardsvacuum com...