Characteristics
Units
3000
1200
1200T
1200TX
600
600T
200
200T
Rated input supply current maximum load
(200 V a.c.)
A (rms)
26
26
26
18
Rated input supply current maximum load
(400 V a.c.)
A (rms)
14
14
14
11
Recommended branch circuit protection UL
(200
‑
230 V a.c. supply)
A
35
35
35
25
Recommended branch circuit protection UL
(380
‑
460 V a.c.)
A
20
20
20
15
Recommended mains conductor size (mini-
mum)
mm
2
(Awg)
6 (10)
6 (10)
6 (10)
6 (10)
Mains connector
-
Han® K
4/4-F
Han® 40A
Note:
Install to local, national, and regional electrical requirements/codes.
Table 8
General electrical data
Description
Units
Rating
Supply voltage 3-phase
V a.c.
200 - 230 or 380 - 460 (see rating plate)
Frequency range
Hz
50/60
Wiring configuration
3 wire plus earth (ground)
Branch circuit protection requirement
Current rating, refer to
Fuse Class gG (IEC 60269), UL class T, class
J or class RK5, Bussmann type JJS or equiv-
alent I2t characteristic rated to 600 V
Voltage tolerance range
± 10%
Installation category
II (IEC 60664)
Input supply voltage unbalance
Should not exceed 2% when assessed over
any one minute period
Short circuit current rating (when installed
with class T or class J fuses)
kA
30
Second protective earth (ground) conductor
Must be installed with cross-sectional area
equal to phase conductor size, up to 16
mm
2
Typical earth leakage*
▪ for 200 - 230 V a.c. systems
mA
< 5
10/2020 - ©Edwards Limited
Page 19
M56635880_D
M56635880_D - Technical data
Summary of Contents for iXM Series
Page 1: ...Dry Pump System iXM INSTRUCTION MANUAL M56635880_D Original instructions edwardsvacuum com...
Page 101: ...This page has been intentionally left blank...
Page 102: ...This page has been intentionally left blank...
Page 103: ...This page has been intentionally left blank...
Page 104: ...edwardsvacuum com...