![Edwards iXM Series Instruction Manual Download Page 76](http://html1.mh-extra.com/html/edwards/ixm-series/ixm-series_instruction-manual_2366129076.webp)
Fault
Warning 63.13 - Sensor Fault
Cause
Sensor is faulty (short circuit)
Remedy
Contact us.
Fault
Warning 70.11 - DP E/C Temp High
Cause
Pump dry end cover temperature is high
Remedy
Check that the cooling water supply is connected and switched on. The supply must
be as per the specification given in
on page 21.
Fault
Warning 70.13 - Sensor missing
Cause
The dry pump end cover temperature sensor has possibly become disconnected
or failed
Remedy
Check that the sensor is installed and is correctly connected.
Fault
Warning 70.13 - Sensor Fault
Cause
Sensor is faulty (short circuit)
Remedy
Contact us.
Fault
Warning 71.13 - AC Sup Missing
Cause
The pump controller cannot communicate with the accessory module
Remedy
1. Check that the accessory module is correctly connected to the pump and
then cycle the power to the pump.
2. If this warning continues, contact us.
Fault
Warning 84.11 - Exh Temp High
Cause
The exhaust pipe temperature is too high.
Remedy
Contact us.
Fault
Warning 84.13 - Sensor missing
Cause
The exhaust pipe temperature sensor has possibly become disconnected or
failed.
Remedy
Contact us.
Fault
Warning 84.13 - Sensor Fault
Cause
Sensor is faulty (short circuit)
Remedy
Contact us.
10/2020 - ©Edwards Limited
Page 73
M56635880_D
M56635880_D - Fault finding
Summary of Contents for iXM Series
Page 1: ...Dry Pump System iXM INSTRUCTION MANUAL M56635880_D Original instructions edwardsvacuum com...
Page 101: ...This page has been intentionally left blank...
Page 102: ...This page has been intentionally left blank...
Page 103: ...This page has been intentionally left blank...
Page 104: ...edwardsvacuum com...