Section 1: Summary of Default Settings
Setup Information
Setup information consists of the process setpoints and the film Processor configurations. All setup
information that was preset at the factory is listed in the tables below.
Setup information is stored in battery backed-up RAM (Random Access Memory). Therefore, you do
not need to program new values every time that you turn on the Processor. Even if power to the
Processor is interrupted or the Processor is turned off, you do not need to program the
Processor.
If you wish, you may change the setup information by using the Soft Keys on the Display Panel. See
the following pages for instructions on how to set up the Processor.
Table 1
Default Processor Configurations for All Cycles
Access Code
4-2-1-3
Processing Cycle
*Standard
Processing Chemicals
RP
Dryer Temperature
*49
°
C (120
°
F)
Time and Date
Operator Must Enter
Pump Calibration
Operator Must Complete
Alarm Volume
Mid Range (7)
Replenishment Mode
Automatic
Temperature Lockout Mode
Disabled
Display Units
˚F, in./min
Receptacle Mode
Safelight Mode
Standby Mode
Interval
Display Language
English
Access to Processor Setup
Allowed
Sleep Mode Options
Roller Jog
Cool Down
Off
Off
Automatic On and Off Timers
Off