Calibrating the Replenishment System
Description:
Calibrate the replenishment system every 3 months.
Calibrating the replenishment system determines the actual rate of the processing solution flowing
through the Replenishment Pumps. To calibrate the Pump you need to first measure the volume of
solution pumped during a set time period, and secondly enter the information into the microprocessor
by using the procedure outlined below. The microprocessor then computes the flow rate of the solution
through the Pump and adjusts the length of time that the Pump must operate so that the volume of
replenishment delivered equals the replenishment volume selected.
Important
The volume of replenishment solution actually measured during this procedure is not the volume
delivered for a 35 x 43 cm sheet of film.
Be sure to do the procedure below twice: once to calibrate the developer replenishment and once to
calibrate the fixer replenishment.
Procedure:
[1] From the Walk-Up Menu, press the “GO TO SETUP” key.
[2] Enter the 4-digit access code.
[3] Press the “MORE” key.
[4] Press the “SETUP” key.
[5] Press the “PUMP CALIB” key.
[6] Press the “DEV CAL” key to calibrate the Developer Pump or the “FIX CAL” key to calibrate the Fixer Pump.