BEFORE EACH USE
• Check the mains lead (must not be pinched or crushed, and no wires must be exposed).
• Check the cable connecting the base to the applicator (for the absence of visible
deterioration).
• Check the external (photo
1
) and internal (photo
2
) optical parts of
.
VERIFICATION OF EXTERNAL OPTICAL PARTS
Check that the glass end of the applicator (optical guide) is free of any surface defects
and, if this is not the case, it is essential that the mobile head should be replaced. This
part can be ordered on the website at
www.e-swin-pro.com
(go to “Accessories and
cartridges” section, then “Mobile applicator head”)or from your authorised distributors.
VERIFICATION OF INTERNAL OPTICAL PARTS
Before each use and before connecting your
, to supply, you must check the
integrity of the internal optical parts: tube,
Filter
(photo
2
) (flat element of red-coloured
glass), optical guide.
The
filter
(flat element of red-coloured glass), must be intact, free of cracks, flat and
regular. It must be clean and dust-free. In order to improve the efficiency of
, it
is recommended that the
filter
should be cleaned before each use, using the
Clean
spray supplied with
. Any
filter
which is damaged as a result of inadequate
maintenance will not be covered by the terms of the guarantee.