Disinfecting/cleaning/sterilising schedule for
Savina Intensive Care Ventilator
Applicable for use with non-infectious patients.
For infectious patients, all parts that conduct breathing gas
must be additionally sterilised after disinfecting and
cleaning.
The parts that conduct breathing gas listed here can be
steam-sterilised at 134
o
C. See "Sterilising" column.
This schedule contains guideline values only. The protocols
of the responsible hospital hygiene official remain
unaffected!
Part
How often
How
Reusable
components
Recommended
cleaning
intervals
Disinfecting and cleaning
Sterilising
Autoclaving at
93
o
C 10 minutes
Wiping
Bath
immersion
Steam
134
o
C
10 minutes
Savina
basic device
per patient
no
outside
no
no
Trolley
Hinged arm
Medical gas hose
per patient
no
outside
no
no
Ventilation hoses,
Y-piece,
Water traps and
Collecting jars
per patient/
weekly
yes
no
possible
yes
Expiration valve
per patient/
weekly
yes
no
possible
yes
Flow sensor
daily
no*
*
Special treatment, bath disinfect in 70 % ethanol, see page 70
outside
possible*
no
Temperature sensor
daily
no
yes
no
yes
Summary of Contents for Savina
Page 1: ...D Dräger Medical Instructions for Use Software 2 n mt78799 Savina Intensive Care Ventilator ...
Page 4: ...4 ...
Page 7: ...7 Intended Use Intended Use 8 ...
Page 16: ...16 ...
Page 46: ...46 ...
Page 54: ...54 ...
Page 55: ...55 Fault Cause Remedy Fault Cause Remedy 56 ...
Page 66: ...66 ...
Page 116: ...116 ...
Page 117: ...117 Technical data Technical data 118 ...
Page 125: ...125 Order List Order List 126 ...
Page 130: ...130 ...
Page 131: ...131 ...