Fault – Cause – Remedy
63
! O
2
monitoring off
O
2
monitoring is switched off.
Switched-off state has been confirmed.
Switch O
2
monitoring on again, see
page 50 or immediately ensure
adequate external monitoring.
!!! O
2
supply down
O
2
supply pressure too low.
Ensure pressure is greater than 2.7 bar.
! O
2
supply down
O
2
supply pressure too low.
O
2
supply pressure is not required
when FiO
2
= 21 Vol.%.
Ensure pressure is greater than 2.7 bar.
!! O
2
supply pressure high
O
2
supply pressure too high.
Make sure pressure is less than 6 bar.
! O
2
supply pressure high
O
2
supply pressure too high.
O
2
supply pressure is not required
when FiO
2
= 21 Vol.%.
Make sure pressure is less than 6 bar.
!!! PEEP high
Expiration tract incorrectly positioned.
Check hose system and expiration valve.
Increased expiratory resistance.
Check bacterial filter.
Replace if necessary.
!!! PEEP inop.
PEEP set not reached.
Check system for leaks.
Ventilation with Savina can be continued,
though the PEEP set will not be reached.
Observe the patient closely.
Call DrägerService.
!!! Pressure meas. inop.
Fluid in expiration valve.
Replace expiration valve,
then disinfect/clean and dry,
see page 71.
Pressure measurement malfunction.
Disconnect patient from Savina and
continue ventilation without delay, using
another independent ventilator. Use
PEEP and/or an increased O
2
concentration if necessary.
Call DrägerService.
!! Rotary knob failed
Rotary knob cannot be turned or pressed. Disconnect patient from Savina and
continue ventilation without delay, using
another independent ventilator. Use
PEEP and/or an increased O
2
concentration if necessary.
Call DrägerService.
!! Rotary knob overused
Rotary knob has been pressed very
frequently within a short time.
Press
»Alarm Reset«
to acknowledge
message.
If this message appears again:
Disconnect patient from Savina and
continue ventilation without delay, using
another independent ventilator. Use
PEEP and/or an increased O
2
concentration if necessary.
Call DrägerService.
Message
Cause
Remedy
Summary of Contents for Savina
Page 1: ...D Dräger Medical Instructions for Use Software 2 n mt78799 Savina Intensive Care Ventilator ...
Page 4: ...4 ...
Page 7: ...7 Intended Use Intended Use 8 ...
Page 16: ...16 ...
Page 46: ...46 ...
Page 54: ...54 ...
Page 55: ...55 Fault Cause Remedy Fault Cause Remedy 56 ...
Page 66: ...66 ...
Page 116: ...116 ...
Page 117: ...117 Technical data Technical data 118 ...
Page 125: ...125 Order List Order List 126 ...
Page 130: ...130 ...
Page 131: ...131 ...