Technical data
120
Flow measurement
(hot-wire sensor)
Inspiratory peak flow FlowPeak
Range
0 to 196 L/min
Resolution
1 L/min
Accuracy
±8 % of measured value
Minute volume MV
Spontaneous breathed minute volume MV
spon
Range
0 to 99 L/min, BTPS*
*
BTPS
Body Temperature, Pressure, Saturated.
Measured values with reference to the conditions of the patient’s lung,
body temperature 37
o
C, ambient pressure, steam-saturated gas.
Resolution
0.1 L/min
Accuracy
±8 % of measured value
T
0...90
approx. 35 s
Leakage minute volume MVleak
based on the inspiratory minute volume
(only in »Mask/NIV« mode)
Range
0 to 100 %
Resolution
1 %
Accuracy
±18 % of measured value
Measured expiratory tidal volume V
Te
Range
0 to 3999 mL, BTPS*
Resolution
1 mL
Accuracy
±8 % of measured value
Leakage-compensated inspiratory measured tidal
volume VTpat
(only in »Mask/NIV« mode)
Range
0 to 3999 mL, BTPS*
Resolution
1 mL
Accuracy
±18 % of measured value
Frequency Measurement
Spontaneous breathing frequency fspn
Total frequency ftot
Range
0 to 150 /min
Resolution
1 /min
Accuracy
±1 /min
T
0...90
approx. 35 s
Inspiration/expiration ratio I:E
Range
0 : 150 to 150 : 1
Resolution
0.1
Accuracy
±6 % of measured value
Inspiration time Tinsp
Range
0 to 15 s
Resolution
0.1 s
Accuracy
0.1 s
Summary of Contents for Savina
Page 1: ...D Dräger Medical Instructions for Use Software 2 n mt78799 Savina Intensive Care Ventilator ...
Page 4: ...4 ...
Page 7: ...7 Intended Use Intended Use 8 ...
Page 16: ...16 ...
Page 46: ...46 ...
Page 54: ...54 ...
Page 55: ...55 Fault Cause Remedy Fault Cause Remedy 56 ...
Page 66: ...66 ...
Page 116: ...116 ...
Page 117: ...117 Technical data Technical data 118 ...
Page 125: ...125 Order List Order List 126 ...
Page 130: ...130 ...
Page 131: ...131 ...