Configuration
53
Configuring the MEDIBUS
*
protocol
The following parameters can be set:
— Baudrate
— Parity check bits
— Number of stop bits
●
Press »
Config.
∆∆
« key repeatedly until
»Configuration 3/4«
appears.
●
Select line for the corresponding interface parameters,
e.g. select baudrate = turn the rotary knob,
to release for setting = press rotary knob.
●
Set value = turn the rotary knob,
confirm = press the rotary knob.
Displaying the device values
●
Press »
Config.
∆∆
« key repeatedly until
»Configuration 4/4«
appears.
Savina displays:
—
SW
Software version of the device
—
Device-ID
specific device identification number
—
Working hours
the total number of hours the device has been operating,
and
—
h since service
the number of working hours since the last inspection
and service by trained service personnel.
●
Release code:
Numerical code for release of options
Return to main page
●
Press the »
Curves
« key. The main page showing a
curve appears.
*
MEDIBUS:
Dräger communications protocol for medical appliances.
083 3
7
388
1
06 3
7
388