![Dräger Savina Instructions For Use Manual Download Page 61](http://html.mh-extra.com/html/drager/savina/savina_instructions-for-use-manual_2526834061.webp)
Fault – Cause – Remedy
61
!! Main switch overused
Savina has detected very frequent
operation of the switch within a short time.
Acknowledge message by pressing
»Alarm Reset«
.
If message appears again:
Ventilation with Savina can be continued.
If Savina cannot be switched off after
ventilation:
Turn the power switch to »
« and press
the »
Standby
« key to switch Savina
off.
Call DrägerService.
!!! Malfunction fan
Fan failure.
Savina uses the fan to cool the interior of
the device and to avoid increased fire risk
in the event of an O
2
leak.
If the »!!! Malfunction fan« alarm message
occurs:
Disconnect patient from Savina and
continue ventilation without delay, using
another independent ventilator. Use
PEEP and/or an increased O
2
concentration if necessary.
Call DrägerService.
! MEDIBUS COM. inop.
The connector of the MEDIBUS cable has
been unplugged during operation.
Reconnect the connector and secure
against disconnection with the two screws.
MEDIBUS cable defective.
Use a new MEDIBUS cable.
Interface defective.
Ventilation with Savina can be continued.
Call DrägerService.
!! Microfilter blocked
Microfilter clogged.
Change microfilter, see page 75.
!!! Microfilter missing
Microfilter is missing or is not
correctly fitted.
Fit microfilter, see page 75.
!!! MV high
The minute volume has exceeded the
upper alarm limit.
Check condition of patient.
Check pattern of ventilation.
Correct alarm limit if necessary.
Flow sensor faulty.
Replace flow sensor.
Water in flow sensor.
Drain water trap,
dry flow sensor.
Machine malfunction.
Disconnect patient from Savina and
continue ventilation without delay, using
another independent ventilator. Use
PEEP and/or an increased O
2
concentration if necessary.
Call DrägerService.
Message
Cause
Remedy
Summary of Contents for Savina
Page 1: ...D Dräger Medical Instructions for Use Software 2 n mt78799 Savina Intensive Care Ventilator ...
Page 4: ...4 ...
Page 7: ...7 Intended Use Intended Use 8 ...
Page 16: ...16 ...
Page 46: ...46 ...
Page 54: ...54 ...
Page 55: ...55 Fault Cause Remedy Fault Cause Remedy 56 ...
Page 66: ...66 ...
Page 116: ...116 ...
Page 117: ...117 Technical data Technical data 118 ...
Page 125: ...125 Order List Order List 126 ...
Page 130: ...130 ...
Page 131: ...131 ...