Operation
34
In »Mask/NIV« mode, another page can be displayed
»
Values 3/3
«.
NIV – Non-invasive ventilation
Mask ventilation
(optional)
Installing the NIV option
This should only be done by experts, following the
corresponding installation instructions.
Using NIV
*
All ventilation modes are selectable in »
Mask/NIV
« mode.
Selecting »Mask/NIV« mode
●
Switch apparatus to standby = hold down »
Standby
«
key for about 3 seconds.
●
Press »
Alarm Reset
« to switch off the audible alarm.
●
Select »
Patient Connection
« line = press rotary knob,
●
Select »
Mask/NIV
« line = press rotary knob,
●
Confirm = press rotary knob.
When masks are used, the dead space increases.
●
Follow the mask manufacturer’s directions.
*
For a detailed description of NIV, see page 106.
Never ventilate an intubated patient in the »Mask/NIV«
mode.
●
After changing over from »Mask/NIV« to »Tube« mode,
check the alarm limits.
●
Avoid high airway pressures.
Risk of aspiration.
1
25 3
7
388
1
26 3
7
388