![Dräger Savina Instructions For Use Manual Download Page 50](http://html.mh-extra.com/html/drager/savina/savina_instructions-for-use-manual_2526834050.webp)
Configuration
50
Message in the information line on screen:
O2 calibration active
After 60 seconds the following appears in the information line:
Reconnect patient
●
Immediately reconnect patient.
Thereafter appears:
O2 calibration ok
The O
2
calibration is completed, Savina ventilates once more
using the original settings.
Disconnection and change of O
2
concentration would activate
the alarms. For this reason Savina turns the alarms off during
calibration.
If the patient is not reconnected within 30 seconds, Savina
continues ventilating and all the alarms are reactivated.
If the following alarm appears after calibration:
!!! O
2
measurement inop.
or
!!! FiO
2
low
●
Replace O
2
sensor, see page 77.
After replacing the O
2
sensor:
●
Allow about 20 minutes warm-up time to elapse,
then calibrate.
FiO
2
- or flow monitoring ON/OFF
For instance, when a used O
2
or flow sensor cannot be
replaced immediately.
●
Use alternative monitoring, as otherwise there will be a
restricted monitoring function.
●
Press »
Config.
∆∆
« key repeatedly until
»Configuration 2/4«
appears.
●
Select the desired line = turn the rotary knob,
confirm = press the rotary knob
●
Set to
»ON«
= turn the rotary knob,
confirm = press the rotary knob.
The corresponding monitoring is activated.
●
Set to
»OFF«
= turn the rotary knob,
confirm = press the rotary knob.
Display in Alarm line:
! Flow monitoring off
or
! O
2
monitoring off
O
2
2
2
2
monitoring and flow monitoring and associated
alarms are permanently deactivated.
●
Provide alternative monitoring.
1
1
5 3
7
388
1
20 3
7
388
080 3
7
388
Summary of Contents for Savina
Page 1: ...D Dräger Medical Instructions for Use Software 2 n mt78799 Savina Intensive Care Ventilator ...
Page 4: ...4 ...
Page 7: ...7 Intended Use Intended Use 8 ...
Page 16: ...16 ...
Page 46: ...46 ...
Page 54: ...54 ...
Page 55: ...55 Fault Cause Remedy Fault Cause Remedy 56 ...
Page 66: ...66 ...
Page 116: ...116 ...
Page 117: ...117 Technical data Technical data 118 ...
Page 125: ...125 Order List Order List 126 ...
Page 130: ...130 ...
Page 131: ...131 ...