Operation
30
Setting Alarm Limits
●
Press »
Alarms
∆∆
« key.
Example display: screen page »
Alarms1/1
«
This page displays all the alarm limits that can be set.
<
= lower alarm limit
>
= upper alarm limit
Example; setting the upper alarm limit for Paw.
●
On the screen, select the line »
Paw
« =
turn the rotary knob,
to release for setting = press rotary knob.
●
Set the desired value = turn the rotary knob,
confirm = press the rotary knob.
The lower alarm limit does not have to be set for the airway
pressure Paw, because it is automatically coupled to the
PEEP setting.
The alarm limits do not have to be set for the O
2
concentration.
These limits are automatically coupled to the O
2
concentration
setting:
Lower alarm limit:
Setting –4 vol. % (at settings up to 60 vol. %)
Setting –6 vol. % (at settings from 60 to 100 vol. %)
Upper alarm limit:
S4 vol. % (at settings up to 60 vol. %)
S6 vol. % (at settings from 60 to 100 vol. %)
A specific screen page »Alarms 2/2« appears in the optional
mode »Mask/NIV«. See "NIV – Non-invasive ventilation",
page 34.
Alarm
Setting Range
Paw
>
10 to 100 mbar
MV
>
2 to 41 L/min
MV
<
0.5 to 40 L/min
T
Apn.
>
15 to 60 sec
f
tot
>
10 to 120 bpm
V
Ti
>
0.06 to 4 L
063 3
7
388
Summary of Contents for Savina
Page 1: ...D Dräger Medical Instructions for Use Software 2 n mt78799 Savina Intensive Care Ventilator ...
Page 4: ...4 ...
Page 7: ...7 Intended Use Intended Use 8 ...
Page 16: ...16 ...
Page 46: ...46 ...
Page 54: ...54 ...
Page 55: ...55 Fault Cause Remedy Fault Cause Remedy 56 ...
Page 66: ...66 ...
Page 116: ...116 ...
Page 117: ...117 Technical data Technical data 118 ...
Page 125: ...125 Order List Order List 126 ...
Page 130: ...130 ...
Page 131: ...131 ...