Operation
35
●
The following advisory message appears briefly:
Setting the ventilation parameters
●
Proceed as in »
Tube
« mode.
In the »Mask/NIV« mode, the »
Tinsp
« key has an extended
function:
In addition to setting the inspiration time in the mandatory
ventilation modes IPPV, SIMV or BIPAP, it is also used to set
the maximum duration of the ASB strokes during pressure-
supported ventilation.
●
Set Tinsp or the maximum duration of the ASB strokes,
0.2 to max. 4 s.
Automatic trigger leakage compensation
Savina compensates for detection of trigger leakage caused by
the patient up to 25 L/min.
Automatic compensation of set tidal volume
Savina compensates up to 100 % of the measured minute
volume for leaks detected.
Monitoring in the »Mask/NIV« mode
To avoid artefacts in the case of very big leaks, the following
alarms may be switched off:
—
MV
minute volume lower alarm limit
—
V
Ti
inspiratory tidal volume upper alarm limit
—
TApnoea
apnoea monitoring upper alarm limit
To switch off (for example
MV
):
●
Reduce »
MV
« until the message appears:
»
MV
off?
Press and turn rotary knob
«.
●
Confirm = press rotary knob.
●
Continue turning the rotary knob until
OFF
appears in the
display.
●
Confirm = press rotary knob.
●
Only switch off alarms if the safety of the patient will
not be compromised by the absence of an alarm.
1
2
7 3
7
388
Summary of Contents for Savina
Page 1: ...D Dräger Medical Instructions for Use Software 2 n mt78799 Savina Intensive Care Ventilator ...
Page 4: ...4 ...
Page 7: ...7 Intended Use Intended Use 8 ...
Page 16: ...16 ...
Page 46: ...46 ...
Page 54: ...54 ...
Page 55: ...55 Fault Cause Remedy Fault Cause Remedy 56 ...
Page 66: ...66 ...
Page 116: ...116 ...
Page 117: ...117 Technical data Technical data 118 ...
Page 125: ...125 Order List Order List 126 ...
Page 130: ...130 ...
Page 131: ...131 ...