Operation
32
Advisory
Low priority message
1
Yellow LED lights up and remains lit.
Advisory messages are marked with one exclamation mark.
Example:
! Flow monitoring off
Savina generates a two-tone sequence that sounds only once.
To remedy the faults, please refer to the "Troubleshooting"
section starting on page 56.
When the fault has been remedied
— the audible alarm is switched off.
— Caution messages (!!) and advisory messages (!) disappear
automatically from the screen.
Warning messages (!!!) remain black on a white background on
the screen and need to be acknowledged:
2
Press the »
Alarm Reset
« key.
3
The message is erased from the screen.
Suppressing the audible alarm
for max. 2 minutes:
4
Press the »
« key. Its yellow indicator LED lights up and
the alarm tone will be muted for about 2 minutes. If the
cause of the alarm has not yet been remedied, the audible
alarm will sound again after about 2 minutes.
If you wish to reactivate the audible alarm before the end of this
2-minute muting period:
4
Press the »
« key again. The yellow LED will now go out.
The message remains on the screen.
The alarm tone sounds only once when a higher priority alarm
occurs during the period, when the audible alarm is muted.
008 3
7
2
6
1
D
3
2
4
1
Summary of Contents for Savina
Page 1: ...D Dräger Medical Instructions for Use Software 2 n mt78799 Savina Intensive Care Ventilator ...
Page 4: ...4 ...
Page 7: ...7 Intended Use Intended Use 8 ...
Page 16: ...16 ...
Page 46: ...46 ...
Page 54: ...54 ...
Page 55: ...55 Fault Cause Remedy Fault Cause Remedy 56 ...
Page 66: ...66 ...
Page 116: ...116 ...
Page 117: ...117 Technical data Technical data 118 ...
Page 125: ...125 Order List Order List 126 ...
Page 130: ...130 ...
Page 131: ...131 ...