Trends
90
Instructions for use Babylog 8000 plus SW 5.n
Displaying trends
The progression of the following measured values
over the past 24 hours is stored in the trend
memory.
1
On the main screen, press the
Trend
key.
An segment of the trend memory is displayed. The
starting time and length of the segment can be
changed.
2
Select the desired measured value by pressing
the
Param
key.
3
Use the
and
keys to select the
length of the segment (max. 24 hours, min.
2 hours). The times displayed indicate the
beginning and end of the segment.
4
Use the
and
keys to shift the
segment:
FiO
2
Inspiratory O
2
concentration
Mean
Mean pressure
MV
Minute volume
C
Dynamic compliance
R
Resistance
RVR
Ratio of respiratory rate to tidal volume
Indicator of chance of success upon
weaning the patient from ventilation
2
1111
2
111
2
A
Located at the left end of the bar:
The segment is located at the beginning of the
trend memory.
2
111
3
B
Located at the right end of the bar:
The segment is located at the end of the trend
memory.
A
B